Filtering samples to on-line analyzers



Use these guidelines to ensure proper filtration and delivery of samples to on-line process analyzers.

Ken Perrotta, Division Engineering Manager 
Allan Fish, Product Manager Parker Hannifin 
Corporation Filtration and Separation Division 
Haverhill, Massachusetts 

Increasingly precise process control strategies, often linked to computer control, have accelerated the use of sophisticated on-line chemical composition analyzers in plant applications. Instruments such as gas and liquid chromatographs, ion chromatographs, laser optic instruments, atomic absorption instruments and specific ion analyzers, which were rarities in laboratories ten years ago, are now found routinely in plant settings. 

While continuing miniaturization and “ruggedization” of the electronics are making the instrumentation circuitry more tolerant of the plant environment, a problem that will not diminish is the level of contamination in plant samples compared to laboratory samples. Factors in plant operation that magnify the difficulty of delivering acceptable samples to on-line process analyzers are: 
 • The sample stream must be processed and fed to the analyzer continuously, rather than the batchwise method permissible in the laboratory. 
 • Test frequency in the plant is far greater than in the laboratory. Where a laboratory analyzer might perform 100 tests a month, an on-line plant analyzer could do 100 tests a day. In the high frequency plant usage, trace contaminants that would not be a problem in the laboratory can rapidly build up and cause instrument failure. 
Filter requirements. 
It is not surprising that contaminants in the plant samples are reported to be the most frequent cause of problems with on-line analyzers. 
While the requirements for an effective filter in the sampling line is generally recognized, it is also important to recognize that it is usually necessary to use a filter specifically designed for sampling applications, rather than trying to make do with a general purpose or homemade filter.
The characteristics that a sample filter should have, in addition to filtering out contaminants, are: 
 • The filter must not change the composition of the sample, other than to remove unwanted impurities. Therefore, the choice of filter media generally is limited to a few chemically inert materials: glass, stainless steel and PTFE. 
 • Since the sample filter often is in a remote or inconvenient location, it must be capable of operating for a reasonable period between scheduled maintenance checks. Even more important, it should not be susceptible to unscheduled problems, such as filter element plugging or rupturing, between regular maintenance checks. 
 • Sample filter maintenance in the field usually is performed under adverse conditions by personnel who are not trained chemists; therefore, the filter should be designed for easy and uncomplicated maintenance. Filter elements should be rugged and not susceptible to handling damage; the unassembled housing should have a minimum number of loose parts, and the housing should be designed so that it is virtually impossible to install a filter element incorrectly. 
 • A filter should introduce minimum lag time into the system. Lag time can be dealt with in the sample system design (slipstream sampling, for example), but sizable dead volume in the filter housing should be avoided. Since large reservoir volume is desirable in many filter applications - such as compressed air or water filters - filters not specifically designed for sampling usually are not suited for analyzers. Requirements for sample filters range so widely that specifying a filter is best done on a case-by-case basis. 
There is, however, one generalization that applies to all sample filter requirements: the filter must be able to separate efficiently a noncontinuous phase contaminant from the continuous sample stream phase. Specifically, the filter must be able to make the following separations, in addition to removing solid particles: • Gas samples - remove liquid droplets 
 • Liquid samples - remove immiscible liquid droplets and gas bubbles. Most filter media will do an adequate job of removing solid particles from liquids or gases, but the only practical commercial media that will separate liquids from gases, gas bubbles from liquids, and two immiscible liquids is resin-bonded glass microfiber media. All recommendations in this paper are based on resin-bonded glass fiber media.


                                Figure 1: External reservoir for high pressure applications



                                Figure 2: Slipstream or bypass filtration



                             Figure 3: Slipstream plus coalescing filtration 

Separating liquids from gases. Glass microfiber filter tubes efficiently separate suspended liquids from
gases. The filter fibers capture the fine liquid droplets suspended in the gas and cause the droplets to run together to form large drops within the depth of the filter tube. 
The large drops are then forced by the gas flow to the downstream surface of the filter tube, from which the liquid drains by gravity. 
This process is called “coalescing”. Since the coalesced liquid drains from the tubes at the same rate that liquid droplets enter the tubes, the tubes have an unlimited life when coalescing liquids from relatively clean gases, and the filters operate at their initial retention efficiency even when wet with liquid. 
The flow direction is inside-tooutside to permit the liquid to drip from the outside of the filter to the housing drain. 
The filter tube grade should be selected for maximum liquid drainage rate, rather than maximum filtration efficiency rating. 
Since liquid drainage rate decreases with increasing filtration efficiency rating, this is a clear case where overspecifying filter efficiency will lead to unsatisfactory results. 
If liquid is carried into the filter in slugs rather than dispersed as droplets in the gas, a filter that is properly sized for steady-state conditions can be flooded and permit liquid carryover. 
If slugging of liquid is expected, a filter with a relatively large bowl should be selected to provide adequate liquid holding capacity, and positive provisions should be made to drain the liquid automatically from the bowl of the housing as fast as it accumulates. 
An automatic float drain can be used if the pressure is in the 10 to 400 psig range. Above 400 psig, the upper limit for commercially-available float drains, the possibilities are: a constant bleed drain, a valve with an automatic timed actuator, or an external reservoir with manual valves (see Figure 1). 
The external reservoir can be constructed of pipe or tubing with sufficient volume to hold all the liquid that is expected to be collected during any period of unattended operation. 
To drain liquid while the filter is operating at pressure or vacuum conditions, valve #1 is closed and valve #2 is opened. If the filter is under vacuum, the external reservoir is a practical method of collecting coalesced liquid for manual draining periodically. 
Alternatively, if any external vacuum source is available, such as an aspirator, the liquid may be drained continuously from the housing drain port. 

Separating two liquid phases

In principle, glass microfiber filter tubes separate suspended droplets of a liquid which is immiscible in another liquid by the same process as they separate droplets of liquid form a gas. 
The liquid droplets suspended in the continuous liquid phase are trapped on the fibers and run together to form large drops, which are then forced through the filter to the downstream surface. 
The large drops separate from the continuous liquid phase by gravity difference, settling if heavier than the continuous phase and rising if lighter. 
The coalescing action of glass microfiber filters is effective with aqueous droplets suspended in oil or other hydrocarbons, and also with oil in water suspensions. 
In practice, however, liquid-liquid separations are much more difficult than liquid-gas separations. The specific gravity difference between two liquids is always less than between a liquid and a gas, and therefore, a longer phase separation time is needed. 
Either the filter housing must be oversized or the flow rate greatly reduced to avoid carryover of the coalesced phase. As a rule of thumb, flow rate for liquidliquid separation should be no more than one-fifth the flow rate for solid-liquid separation. 

Even at low flow rates, if the specific gravity difference between the two liquids is less than 0.1 units (for example, if an oil suspended in water has a specificgravity between the two liquids is less than 0.1 units (for example, if an oil suspended in water has a specific gravity between 0.9 and 1.1), the separation time for the coalesced phase may be impractically long.

In that case, if there is only a small quantity of suspended liquid, the filter tube can be used until saturated with the suspended liquid and then changed.

Another practical problem with liquid-liquid separations is that small quantities of impurities can act as surface-active agents and interfere with the coalescing action. For that reason it is not possible to predict accurately the performance of a liquid-liquid coalescing filter, and each system must be tested on site. 

The general guidelines for the system to start testing are to use 25 micron filter tubes and flow inside-out at very low flow rates. If the suspended liquid is lighter than the continuous phase the housing should be oriented so that the drain port is up. 

Removing gas bubbles from liquids. Glass microfiber filter tubes readily remove suspended gas bubbles from liquid, eliminating the need for de aeration tanks, baffles or other separation devices. Flow direction through the filter is outside-to-inside, and the separated gas bubbles rise to the top of the housing and are vented through the drain port.

If slipstream sampling is used (see below), the separated bubbles are swept out of the housing with the bypassed liquid. Filter tubes rated at 25 micron are a good choice for gas bubble separation. 

Slipstream or bypass sampling.  Instrument sample use rates are invariably quite low, yet it is essential to minimize lag time in the sample system. Since analyzers often are located some distance from the sampling point, samples usually are transported to the analyzer at a relatively high flow rate to minimize lag time. The sample is divided at the analyzer, with the analyzer using the portion it requires (usually a very small fraction of the total sample), and the balance recycled to the process or vented.

If the sample filter is located in the low-flow line to the analyzer, it will have good life between filter element changes because the solids loading rate is very low; however, the filter must be carefully selected to avoid introducing unacceptable lag time.

If the filter is located in the high-flow portion of the sample system, its effect on sample lag time can be relatively low, but the life between filter changes may be inconveniently short because the element is filtering a much greater volume of material than the analyzer is using.

Ideally, a filter should be located at the point where the low-flow stream is withdrawn to the analyzer (see Figure 2). This arrangement permits the main volume of the filter to be swept continuously by the high flow rate system, thus minimizing lag time.

At the same time, only the low-flow stream to the analyzer is filtered, thus maximizing filter life. A slipstream filter requires inlet and outlet ports at opposite ends of the filter element to allow the high flow rate of the by-passed material to sweep the surface of the filter element and the filter reservoir, and a third port connected to the low flow rate line to the analyzer, which allows filtered samples to be withdrawn from the filter reservoir.

If bubble removal from a liquid is a requirement; this function may be combined with slipstream filtration, since the recommended flow direction for bubble removal is outside-to-inside, and the separated bubbles will be swept out of the housing by the bypass stream. In this case the liquid feed should enter at the bottom of the housing and the bypass liquid exit at the top of the housing. A special problem arises in slipstream




                                                   Figure 4: Stack gas sampling     
                                                 

                                                      
Figure 5: High pressure steam sampling

Hydrophobic membrane sampling. 

Many online instruments are susceptible to corrosion and skewed analysis from any water and moisture contamination. Gas chromatographs, mass spectrometers, oxygen analyzers and other sensitive on-line instruments require complete removal of all moisture.

Instrument sensitivity levels range from PPM to PPB and “percent level” component concentrations. As a result, it is good practice to install a hydrophobic membrane filter in line to the instrument to protect and safeguard it from any moisture contamination.

A stainless steel filter housing with a hydrophobic membrane allows the sample gas to flow on the upstream side of the membrane and exit through the outlet port on the downstream side (see Figure 6 



                Fig.6 - Stainless steel filter housing with a hydrophobic membrane

Slipstream or bypass filtration 

Entrained moisture will not flow through the membrane and will exit out the by-pass port on the upstream side of the membrane completely protecting the instrument from moisture. If the sample gas contains excessive amounts of moisture and particulates it is recommended to use a stainless steel filter housing that incorporates a coalescing prefilter and the hydrophobic membrane (see Figure 7). The coalescing prefilter will protect the membrane from premature blinding and extend its useful life.



                       Fig.7 Stainless steel filter housing that incorporates a coalescing prefilter and the                                        hydrophobic membrane

Minimal panel space, permanent line mount sampling. As more and more sample lines and instruments are added to instrument sheds, the need to utilize the space in the most efficient way has become critical. In addition, the need to maintain sample filters without having to “break” the sample line and expose it to ambient conditions has also become quite important to most facilities. By not having to “break” sample line the need to flush the lines prior to start up is eliminated. 

For this application, it is recommended a stainless steel filter designed to be horizontally mounted at a 10 degree angle be installed to minimize the amount of foot print on the panel (see Figure 8).


                       Fig.8 Stainless steel filter designed to be horizontally mounted 

This design incorporates the inlet ,outlet and drain ports all in the head of the filter enabling filter element change outs without disrupting the sample line.